Blood Bank Refrigerators are designed in accordance with FDA listed Class II medical devices. In addition, blood bank refrigerators also conform to the requirements set forth by AABB for the refrigerated cold storage of blood-based products. Backed by optimal temperature control and EPA SNAP compliant refrigerants, these high-performance units protect blood, prevent waste, and allow for peak delivery. Blood bank refrigerators utilize smart controllers and feature a full array of alarms, LED interior lighting, stainless steel interiors, sliding drawers and probe access port.
Environmental
We are an industry leader in providing a variety of Refrigerators and Freezers that are built with Hydrocarbon Refrigerants. Environmentally friendly, natural hydrocarbon (HC) refrigerants (not containing hydrofluorocarbons) vastly reduce global-warming potential (GWP) while meeting new EPA/SNAP mandates and UL, ASHRAE compliance guidelines.
Storage Capacity: 23 Cu. Ft.
Door: Glass, right hinged, self closing, magnetic door gasket for positive seal
Handle(s): Heavy Duty
Drawers: 7 drawers standard with option for 8th, 304 stainless steel drawers, 65 lb. capacity each
Interior Lighting: LED Light
External Probe Access: Probe access port (3/4")
Insulation: High density urethane foam cabinet and door insulation
Exterior Material: Powder coat steel
Access Control: Keyed door lock
General Warranty: 2 year parts and labor (excluding calibrations), 7 year compressor warranty
Weight(lbs.): 414 lbs.
Shipping Weight(lbs.): 473 lbs.
Rated Amperage: 3
Power Plug/Cord: NEMA-5-15P
Compressor: Top mounted, variable speed compressor
Refrigerant: R290
Defrost: Cycle optimized, zero energy
Uniformity (Cabinet air): '+/- 0.7°C
Stability (Cabinet air): '+/- 0.5°C
Maximum temperature variation: '+/- 0.8°C
Uniformity (Simulator ballast): '+/- 0.1°C
Stability (Simulator bag): '+/- 0.1°C
Controller technology: Proportional Integral Derivative (PID) microprocessor with LCD display, °C/°F switchable
Battery Backup: Controller
Digital Communication: RS-485 (MODBUS)
Temperature setpoint range: 1°C to 10°C (34°F to 50°F)
Display probe: 2 Stainless steel calibrated probes. Upper probe: 4 oz. (120 ml) bottle, 50% glycerol mixture. Lower probe: Solid thermal media
External alarm connection: State switching State switching remote alarm contacts
Alarms: Visual and audible indicators, power failure, temperature sensor failure, battery voltage monitor and replacement, high/low temperature, door ajar